Skip to content

Commit 909e98f

Browse files
cryogenics/Oxford-Instruments-IPS.md: Corrected table formatting.
1 parent 295f303 commit 909e98f

File tree

1 file changed

+5
-5
lines changed

1 file changed

+5
-5
lines changed

doc/specific_iocs/cryogenics/Oxford-Instruments-IPS.md

Lines changed: 5 additions & 5 deletions
Original file line numberDiff line numberDiff line change
@@ -110,11 +110,11 @@ stateDiagram-v2
110110

111111
There are various IPS units here is a list of the ones we know the settings for (Lab view settings are in `...LabVIEW Modules\Oxford Software\TESLTRON.INI`):
112112

113-
Setting | Description | IN LabVIEW | 7T Magnet | 9T Chopper Magnet (LET)
114-
------- | ----------- |---------------------------------------------|------------ | ---------------
115-
`STABILITY_VOLTAGE` | 5 voltages are compared and they must differ by less than this to be declared stable | hard coded | 0.1 | 0.1
116-
`MAX_FIELD` | Maximum field | `PS0MaxField(2.2K)` and `PS0MaxField(4.2K)` | 7.0 [1] | 8.8 [2]
117-
`MAX_SWEEP_RATE` | Maximum rate of change of the field in B/min | `PS0MaxSweep(4.2K)` and `PS0MaxSweep(2.2K)` | 0.399 | 0.3990
113+
| Setting | Description | IN LabVIEW | 7T Magnet | 9T Chopper Magnet (LET)|
114+
|---------------------| ----------- |---------------------------------------------|------------ | ---------------|
115+
| `STABILITY_VOLTAGE` | 5 voltages are compared and they must differ by less than this to be declared stable | hard coded | 0.1 | 0.1|
116+
| `MAX_FIELD` | Maximum field | `PS0MaxField(2.2K)` and `PS0MaxField(4.2K)` | 7.0 [1] | 8.8 [2]|
117+
| `MAX_SWEEP_RATE` | Maximum rate of change of the field in B/min | `PS0MaxSweep(4.2K)` and `PS0MaxSweep(2.2K)` | 0.399 | 0.3990|
118118

119119
[1] - The limit may be lower on some beamlines due to the construction of their sample area. Below are known cases: